Removal of transverse carriage and patient table 

Contents 



Siemens; AG 

Medical LiKjineerniy Group 


RLL5-310.091.03.01.02 
Rev. 01 



Removal and installation of the transverse carriage 

Safety note, documents required, tools required 


1 


Safety note 

Note! When performing work and tests, observe the product-specific safety notes 
contained in this document and the general safety notes in register 2 of 
the Tl folder. 

Tests and settings which need to be made under radiation are marked with the ra¬ 
diation symbol V . During all procedures marked with this symbol, radiation 
protective precautions must be observed strictly. 


Documents required 

Operating instructions, UROSKOP D... 
Wiring diagram 


Tools required 

Service tools 


Siemens AG 
Medizinische Technik 


RLL5-310.091.03.01.01 
Rev.01 




2 


Removal and installation of the transverse carriage 

Safety note, documents required, tools required 



Siemens AG 
Medizinische Technik 


RLL5-310.091.03.01.01 
Rev.01 




Siemens AG 

Medical Engineering Group 


RLL5-310.091.03.01.02 
Rev.01 












4 Removal and installation of the transverse carriage 

Procedure 



• Unscrew the cover plate (C/Fig.5) and the cover housing (D/Fig.5). 

• Undo the plug connection M7.X1 (Fig.6) and unscrew the energy chain 
(2 Allen screws, VFig.6). 

• Unscrew the right head-end toothed rack 

(4 hexagonal screws, 2 behind the motor and 2 see /'/Fig.6). 

• Unscrew the left toothed rack (4 Allen screws /'/Fig.7). 

• Unscrew the table transverse carriage 

(6 hexagonal screws, 2 behind the motor and 4 see /'/Fig.8). 

• Pull the table transverse carriage all the way to the front with two people. 

Note: After the transverse carriage has been pulled out, do not move the rail on the 
unit under any circumstances, since the balls could fall out of the linear ball 
bearing 

• Reinstall the tranverse carriage in reverse order. 


TDRL 2/Wendt, TDRL 2/Schiee 


Siemens AG 

Medical Engineering Group 


RLL5-310.091.03.01.02 
Rev.01