Spirolab II 

Service Manual 


Date 

Date 



Service Manual Rev. 0 

Prepared by Arne Langenkamp 
06/04/01 

Approved by Paolo Sacco Boschetti 
06/04/01 



MIR srl 


Spirolab / Spirolab II - Service Manual 


1. INTRODUCTION 

1.1. INTRODUCTORY REMARKS 

1.1.1. General Aspects 

The spirometers from the product group MIR 010 are marketed under the labels 

spirolab and spirolab II. 

The main difference between the 2 models spirolab and spirolab II is principally an 
LCD display in colour in the model spirolab II while the model spirolab has a black 
and white LCD display. 

From here onwards for the sake of brevity, where not indicated differently with the 
wording spirolab we refer to all models of the product group. 

This manual has been written for the technicians involved in the service of the spiro¬ 
lab. Service can be carried out by the service organization of the manufacturer or by 
any other technician authorized by MIR srl. 

The manufacturer will not be held responsible for the results of maintenance or repairs 
by unauthorized persons. 

1.1.2. Safety Aspects 

Safe execution of the procedures in this manual requires technical experience and 
general knowledge of safety precautions. 

Before you start any operation on the unit, you must 
have first read both the User’s and this Service manual. 

Warnings, Cautions and Notes are used throughout this manual to bring special 
matters to the immediate attention of the reader. 

The Warning concerns danger to the safety of the technician or user. 

The Caution draws attention to an action which may damage the equipment. 

The Note points out a statement deserving more emphasis than the general text, but 
which does not deserve a Warning or Caution. 

Warning: 

In order to prevent unforeseen incidents, keep away from sensitive electronical 
devices. Inside the spirolab there are high voltages but at very low currents, so 
these represent no danger for persons with experience and with a general 
knowledge of safety precautions. 

In exceptional cases these high voltage may cause interference with other 
electronic devices. 

The sequence of steps in a procedure may also be important from the point of view of 
personal safety and prevention of damage, therefore never change the sequence of 
steps or alter any procedure. 


Page 2 of33 




MIR srl 


Spirolab / Spirolab II - Service Manual 


Warning: 

Batteries may explode if defective or damaged or disposed of in fire. 

Do not short-circuit the batteries! 

1.1.3. Legal Aspects 

This manual and the information herein are property of MIR srl (Rome, Italy). 

Copying this manual in whole or in part is strictly forbidden. 

MIR srl disclaims any responsibility for personal injury and/or damage to equipment 
caused by: 

negligence or disregard of a Warning or Caution 
deviation from any of the prescribed procedures 
execution of activities which are not prescribed 

ignorance of the general safety precautions for handling tools and the use of 
electricity 

incorrect or unreasonable use 

1.1.4. Environmental Aspects 

Please dispose properly of any items replaced, following all the local regulations. 

All components (housing, PCBs, batteries, printer, electronic and disposable para.ts, 
etc.) must be disposed of according to the relevant local regulations. 

1.2. DATA REGISTRATION 

1.2.1. Equipment and customer registration / Service file 

From the point of view of safety and product liability the following data must be re¬ 
gistered for each unit: 

- Equipment data: model and serial number 

- Customer data: date of purchase, name and full address of the customer 

- Service data: log of all service activities 

- Configuration data: the actual configuration of the equipment (hardware and 
software versions). 

The distributor (through their dealer and/or service organization, if any) is responsible 
for maintaining these record and they must be able to provide these data when reques¬ 
ted. 

1.2.2. Configuration register 

A configuration file gives the serial numbers and/or batch numbers of main sub- 
assemblies and/or important para.ts of each unit. The para.ts listed in a configuration 
file are defined as registered para.ts. For example para.ts as mainboards, keyboards, 
battery packs and internal software. 

Note: MIR srl has the original configuration file of each unit sold. Any possible 
changes in the original configuration of the supplied units must be filed and the 
distributor must be able to provide this data. These changes can simply be stored in 


Page 3 of33 




MIR srl 


Spirolab / Spirolab II - Service Manual 


the service file of the specific unit. Para.t, serial or batch number(s) of the new para.t 
must be documented. 

1.2.3. Product documentation 

The documentation set for the spirolab includes a User’s Manual. The User’s Manual 
is a recommended item for all service engineers. 

Ordering data: 

- 980060 User’s Manual 

- 980070 Service manual 

1.2.4. Installation 
Warning: 

The mains connection must comply with the national requirements regarding 
medical facilities. 

Before connecting this appara.atus to the mains supply, check that the mains 
voltage and frequency stated on the type plate of the battery charger 
correspond to that of the local mains supply. The 220-240V model is intended 
for nominal mains voltages of 220, 230 or 240 volt. The 115V model is 
intended for nominal mains voltages of 110 or 120 volt. 

1.2.5. Technical modifications 

This publication could include technical inaccuracies or typographical errors. Changes 
are periodically made to the information herein; these changes will be published as 
Service Info in appendix B of this Service Manual. 

Note: Some functions described in this manual may not be activated in the present 
software or hardware-version. For detailed information, please refer to the User’s 
Manual. 


1.2.6. Product evaluation 

In order continuously to improve the quality of our products, MIR srl requests that the 
service organizations should report any device problem which may be discovered. 
When the same problem regularly occurs or when a safety problems occur, MIR srl 
would like you to send us a product complaint report. The report must contain the 
name, the type of the product, a detailed description of the problem, the frequency of 
the same problems, quantity of sold equipment and your solution to the problem (if 
any). 


Page 4 of33 




MIR srl 


Spirolab / Spirolab II - Service Manual 


1.2.7. Additional Information 

Please do not hesitate to contact us if you require additional information. 

MIR srl 

Via del Maggiolino, 125 

00155 Rome 

Italy 

Tel.: +39/06/2275 4777 
Fax.: +39/06/2275 4785 
Email: mir@spirometry.com 


Page 5 of33 




MIR srl 


Spirolab / Spirolab II - Service Manual 


1.3. GENERAL 

The spirolab is an “open circuit system” for the measurement of inspiratory and 
expiratory lung function para.ameters. It is suitable for basic lung function analysis of 
the mechanical respiratory tract para.ameters. Three different respiratory manoeuvres 
can be performed: the forced vital capacity test (FVC), the slow Vital Capacity test 
(VC/IVC) and the Maximum Volontary Ventilation test (MW). 

The spirolab has been designed and manufactured to ensure the highest level of 
safety and the unit fully complies with the stringent international IEC 601-1 and IEC 
601-1-2 standards. 

1.3.1. Technical Data 

1.3.1.1. Battery charger 

Battery charger supply : 115 VAC, 220-240 VAC 
Permissible mains voltage var. : ±15% (115 V), ± 10% (220-240 V) 

Mains frequency : 50 - 60 Hz 

Max. current : 1 A at 12 V 

1.3.1.2. Spirolab unit 

Max. current inside the unit : 3,2 A (during printing) 

Internal fuses : T 2 A (FUS1) SMD (battery charge controller) 

: T 4 A (FUS2) SMD (main fuse) 

Internal batteries : Ni-MH for main power supply (6x1,2 V), 4 Ah 

: Lithium for data backup (1x3,6 V), 370 mAh 
Approval : CE 0476 EC mark for Medical Devices 
Dimensions : 310x200x65mm 

Weight : 1,9 kg (including battery pack) 

Volume range :10L, BTPS 

Flow range : ± 16 L/s, BTPS 

Volume accuracy : ± 3% or 50 mL 

Flow accuracy : ± 5% or 200 mL/s 

Dynamic resistance at 12 L/s : < 0,8 cmFEO/L/s 

Type of electrical protection: Class 2 appara.atus 

Level of protection against direct or indirect contact: 

Category BF appara.atus 

Level of water ingress protection: Common appara.atus 

Level of assurance of use in the presence of an anaesthetic mix flammable with air 
or oxygen or protoxide of nitrogen 

: Appara.atus not suitable 

Conditions of use : Appara.atus for repeated use 

Storage conditions : Temperature: MIN 0 °C, MAX +40 °C 

Humidity: MIN 10% RH; MAX 95%RH 
Operating conditions : Temperature: MIN +10 °C, MAX +40 °C; 

Humidity: MIN 10% RH; MAX 95%RH 


Page 6 of33 





MIR srl 


Spirolab / Spirolab II - Service Manual 


1.4. STANDARDS APPLIED 

According to rule 10 of the European directive 93/42/CEE the device is classified 
as Class Ha 

EN 60601-1 Electronic medical appara.atus. Para.t 1: General safety standards 
EN 60601-1-1 Electronic Medical appara.atus. Para.t 1: General safety standards 

1. Collateral standard: Safety requirements for medical electronic systems 

EN 60601-1-2 Electronic medical appara.atus. Para.t 1: General safety standards 

2. Collateral standard: Electromagnetic compatibility - Requirements and tests 
EN 60601-1-4 Electronic medical appara.atus. Para.t 1: General safety standards 
4. Collateral standards: Programmable medical electronic systems 

ERS: The European Respiratory Journal - Standardised Lung Function Testing - 
Volume 6 Supplement 16, March 1993 

ATS: Standardisation of Spirometry - 1987 Update - American Review of 
Respiratory Disease 136:pp.1285-1298,1987 

ATS: Standardisation of Spirometry - 1994 Update - American Review of Crit. 
Care 152: pp.1107-1136,1995 


Page 7 of33 




MIR srl 


Spirolab / Spirolab II - Service Manual 


2. HARDWARE DESCRIPTION 

Because of the modular design of the spirolab, the description is on a block diagram 
level. 

The spirolab consists of six modules: 

- Mainboard module 

- Keyboard module 

- Display module 

- Printer module 

- External turbine flowmeter 

- External battery charger 

BLOCK DIAGRAM 


TURBINE FLOWMETER 

Optoelectronic head 
Turbine 

Mouthpiece adaptor 


12 V-BATTERY CHARGER 


DISPLAY MODULE 


MAINBOARD MODULE 

Batteries 
Charging controller 
Power supply circuit 
► Microcontroller 
FLASH memory 
SRAM memoiy 
RTC (Real Time Clock) 
Printer drivers 
Display drivers 
Measuring controller 
Room temp.sensor 
RS 232-port 


PRINTER MODULE 


KEYBOARD MODULE 


Page 8 of33 














MIR srl 


Spirolab / Spirolab II - Service Manual 


2 . 1 . 

2.2. MAINBOARD MODULE 

The mainboard module translates the input signal from the turbine flowmeter into 
spirometry values and compara.es them with the predicted values calculated with the 
para.ameters age, sex, height, weight and ethnic group. Test results are displayed on a 
LCD and can be printed on paper. Spirometry test data are stored into memory and are 
available for later use. 

The mainboard module includes: 

Battery pack (6x1,2V NiMH) for device supply 
Lithium battery 3.6V for data backup (SRAM and RTC) 

Charging controller for battery pack 

DC power supply controlling circuit 

Microcontroller, including EEPROM with “boot” program 

FLASH memory with “boot” and application software 

SRAM memory with device configuration and spirometry data 

RTC real time clock 

Thermal Printer driver 

Graphic LCD driver 

Measuring controller for flow, volume and ambient temperature 

Ambient temperature sensor (to enable conversion from ATP to BTPS 

conditions). 

Serial RS 232 bidirectional communication port with optoisolation (4 KV) for 
connection to a PC, external printer, modem etc. 

2.2.1. Jumpers 

On the mainboard are four jumpers, the position of these jumpers is of crucial 
importance for the correct functioning of the spirolab. 

SJ1 connects the Lithium backup battery with the mainboard. Removal will cause a 
loss of all data, both configuration and test data. 

SJ2 connects the NiMH-battery with the mainboard. If this jumper is removed the unit 
is disconnected from the battery power supply. 

JP1 enables you, by setting a jumper, to program the EEPROM inside the 
microprocessor via the serial RS232 port 

JP2 enables you to switch between three modes to terminate the charging process. 
These are further explained in the next para.agraph. 

Only in the case of a replacement procedure then removal of all jumpers is 
recommended to avoid unforeseen damage to the mainboard and/or to components. 

2.2.2. Charging controller for battery pack 

The charging controlling circuit used in the the spirolab ensures a fast charge and 
optimum condition of the battery pack providing that the battery temperature and 
voltage are within the preset limits. Temperature, voltage and time are all monitored 
throughout the fast charge process. 


Page 9 of33 




MIR srl 


Spirolab / Spirolab II - Service Manual 


The charging process itself is automatically initiated in two situations: 

1. After connecting the battery charger to the unit 

2. When the unit is switched on, the battery charger is connected and the voltage 
level of the battery is below a preset limit. In this situation the LCD indicates the 
low battery status with two red lines. 

The fast charging process is terminated by any of the following: 

Battery voltage out of range (Maximum/Minimum) 

Battery temperature out of range (Maximum/Minimum) 

Maximum charging time (timeout = 5h:30min) 

Termination according to the charging mode, chosen by jumper JP2. 

Charging termination modes: 

1. Temperature Arise Detection (AT/At), JP2 jumper in the default position as in¬ 
dicated on the attached drawing. The battery temperature increases more quickly 
when the battery is fully charged. If the speed of temperature increase inside the 
battery pack exceeds a certain value then the charging process is terminated. The 
charging controller (IC15) samples the thermistor voltage (NTC) every 34 sec., 
and compare it to the value measured two samples earlier. If the voltage has fallen 
16mV ± 4mV or more, the fast charge is terminated. 

2. Peak Voltage Detection (PVD), JP2 jumper in the other possible position: The 
battery’s voltage decreases slightly when it is fully charged. In this charging- 
modes, the microcontroller checks the battery voltage and terminates the charging 
process in the event that the voltage on the battery pack decreases by more than 
the preset limit, that limit is configured by the position of the jumper: In this 
mode, the battery’s voltage is the criteria for terminating the charging process. If 
the voltage decreases by 36mV ± 4mV in 34 seconds this means that the battery is 
fully charged, and the controller then terminates the charging process. 

3. Decrease Voltage Detection (-AV), no JP2 jumper. This mode is correspondent 
to the PVD-mode, but at higher sensibility, because the charging process ends at a 
voltage decrease of 18mV ± 2mV in 34 seconds. 

Caution: 

In case of needing high power while charging (i.e. when printing), the battery’s 

voltage decreases. In this case using the modes Peak voltage Detection or -AV the 

charging process will be terminated, although the battery is not fully charged. 

The factory setting is designed for the AT/At Termination mode and is not to be 

changed to avoid the above-mentioned problems! 

Charging phases: 

1. After connecting the battery charger to the unit, the red LED (CHARGELED) 
starts flashing for few seconds. In this period, the battery is charged in pulses at a 
low frequency and the battery’s response is monitored. If the red LED continues 
to flash this shows that the battery’s temperature is out of range or the battery 
voltage is too low. In this case, the controller enters the charge pending state and 
waits for both conditions to fall (temperature) or rise (voltage) within their al¬ 
lowed limits. There is no time limit on the charge pending state; the charger re¬ 
mains in this state for as long as the voltage or temperature conditions are outside 


Page 10 of 33 




MIR srl 


Spirolab / Spirolab II - Service Manual 


of the allowed limits. If the voltage is too high, the controller goes to the battery 
absent state and waits until a new charge cycle is started. 

2. If everything is in order (battery temperature, temperature rising, battery voltage 
etc.) then the fast charging process starts, the red LED lights continuously. In the 
following 10 minutes, the battery is charged at a lower rate to initialize slowly the 
chemical reaction inside the battery pack. After about 10 minutes, high-frequent 
pulses at the full current rate of 1A charge the battery. 

3. At the end of the regular charging process, the green LED (DONE LED) lights 
continuously. Now it can be assumed that the battery is full charged. 

4. If after completing the charging process the battery charger is still connected, a so- 
called Pulse-Trickle-Function is activated that gives a very low charge to com¬ 
pensate the self-discharging of the battery while it is idle connected to the charger. 

2.2.3. Room temperature sensor 

Transducer of temperature/voltage (IC 12) NATIONAL LM 35 to measure the ambi¬ 
ent (room) temperature to enable the calculation of the BTPS conversion factor. 

The output signal, amplified through an operational amplifier (IC 13, 100 mV/°C) is 
connected to one of the A/D channels of the microprocessor (IC 1, line ROOMT, pin 
35). 

2.2.4. RS 232 communication port 

Bidirectional serial port, controlled by the microprocessor via its serial port. Towards 
the PC the port has 9 lines, configured according to the EIA RS232 Standard: 


Pin 

Description 


1 . 

Received Line Detector 


2. 

Receive Data 

RX 

3. 

Transmit Data 

TX 

4. 

Data Terminal Ready 

DTR 

5. 

Signal Ground 

GND 

6. 

(NC) 


7. 

Request To Send 

RTS 

8. 

(NC) 


9. 

(NC) 



NC = Not connected 

The voltage of the signals according to this standard is in the range of +3 to +12 V for 
High Level (true value) and -3 to -12 V for Low Level (false value). 

Note: Because spirolab is isolated from the connected device by optocouplers, there 
are two electrical grounds. For internal measurements (before the signal reaches the 
optocouplers), there is an internal ground i.e. the negative pole of the battery pack. 
For measurements between the optocouplers and the connected device, you must use 
the external ground, which is different (i.e. pin 5 of CN1). 

2.3. DISPLAY MODULE 


Page 11 of 33 




MIR srl 


Spirolab / Spirolab II - Service Manual 


The display module displays patient data, user set para.ameters and test results. 

It is attached to the body with a hinge and allows an adjustment in angle in a range up 
to 160° with respect to the chassis. 

The display-version depends on the purchased model: 

Graphic B/W type FSTN LCD 320x240 pixel (spirolab) 

Graphic 16-colour type FSTN LCD 320x240 pixel (spirolab II) 

It is connected to the mainboard with a 16-pin flat-cable, while the backlight supply is 
provided by a 2-wire-cable. 

2.4. PRINTER MODULE 

The printer module carries out a printout of the patient data and spirometric results. 
The printer is a thermal printer and requires thermal printing paper. 

The printer is connected to the mainboard via 3 cables: 

a 4-pin flat-cable for driving the motor for paper advance 
a 4-pin flat-cable for the sensor, that checks that paper is inserted 
28-pin flat-cable for supply, data transmission and checking the temperature of the 
printer head, to avoid overheating 

2.5. KEYBOARD MODULE 

The keyboard module constitutes the interface between user and unit. It is connected 
to the mainboard via two 15-pin flat-cables (CN 8, CN 9). 

When the unit is switched off, the only active key is the red ON_OFF key, which 
allows the unit to be switched ON or OFF manually. 

2.6. TURBINE FLOWMETER 

The sensor for measuring flow and volume is the same one already used in other 
spirometers produced by MIR (Series MIR 001 Mod. spirobank and spirobank G). 

The turbine flowmeter consists of three mechanical and two electrical para.ts: 
Mechanical parts: 

optoelectronic head with IR-sensor 
turbine 

mouthpiece adaptor 
Electrical parts: 

Two pairs of infrared transmitters/receivers positioned as shown below. 

A signal conditioning circuit to rectify the output signal from the two infrared 
light receivers. 

The rotation of the rotor causes the interruption of the infrared beam, thus creating a 
pulse signal which has a frequency directly proportional to the flow. 

The measurement of the air flow which passes through the tube is therefore propor¬ 
tional to the number of interruptions of the infrared beam. 


Page 12 of 33 




MIR srl 


Spirolab / Spirolab II - Service Manual 



The phase difference Acp (positive or negative) between the signals from each of the 
two infrared receivers (IR 1 and IR 2) depends upon the direction of rotation of the 
moving rotor and therefore supply the information of the direction of the air flow 
(expiration or inspiration). 

In detail, Acp > 0 for expiratory flow, Acp < 0 for inspiratory flow. 


Page 13 of 33 




MIR srl 


Spirolab / Spirolab II - Service Manual 


1" PULSE 


2° PULSE 


AT 


Acp <0 => Inspiratory flow 


1” PULSE 


2“ PULSE 


| | Acp > 0 => Expiratory flow 

K 

FLOW = - 

AT 

The two pulse trains are squared by a Schmitt trigger (IC 14) and then sent to two 
input ports of the microprocessor (IC1, pins 83 and 84). The microprocessor has the 
possibility to switch the turbine on or off (IC1 pin 78 and Ql). 


Page 14 of 33 




MIR srl 


Spirolab / Spirolab II - Service Manual 


3. MAINTENANCE 

3.1. GENERAL 

We recommend checking the spirolab on an annual basis. 

For cleaning of the spirolab and the accessories please see the User’s Manual. 

3.2. TEST EQUIPMENT 

For the repair and maintenance procedures of the spirolab the following test 
equipment and accessoires are required: 

Complete set of precision engineering tools (including 2.5 mm alien key and 
cross-screwdriver) 

Calibration syringe (3L is recommended) 

Digital multimeter, at least 3'A digits, accuracy better than 1%. 

Oscilloscope 
Digital Thermometer 
- PC 

If a problem cannot be solved with the present equipment and suitable instruments are 
not available, please send the spirolab to the manufacturer to carry out the service. 

3.3. CHECKLIST 

The following procedures must be carried out during an inspection and/or after every 
repair: 

• Modifications, if any, must be noted on the the “Service Info” sheets. See 
appendix B of this manual for further information. 

• Visual inspection. Check the electrical connection for safety and check that all 
components are properly fastened. 

• Check that all accessories do not present any visible damage. 

• Functional test (see. Chapter 4.3) 

• File a copy of all service activities. 

3.3.1. Functional test 

3.3.1.1. Self Test 

Switch on the equipment. The spirolab will carry out the so-called self-test for approx 
3 seconds. It is assumed that when the self-test is passed all functions of the spirolab 
are okay. 

Any severe malfunction (if any) will be reported on the display or the printer. 

Switch off the spirolab. 


Page 15 of 33 




MIR srl 


Spirolab / Spirolab II - Service Manual 


3.3.I.2. Software version 

The software version is indicated in the upper line of the main data screen and on 
every printout. 


3.3.I.3. Hardware control 

All these points must be controlled before doing any measuring test or replacement on 
the unit. 

1. The housing does not show any severe damages like scratches, splits or breakages, 
that could endanger a secure use. 

2. All 5 rubber feet on the chassis are present and not damaged. 

3. All pixels on the display are visible and there are no scratches on the display. 

4. Time and date are correct. 

5. Pressing any key (except the ON/OFF-Key and the keys to adjust the display) 
causes a short beep. 

6. The PCB is firmly attached to the chassis by 8 screws. 

7. The printer is firmly attached to the chassis by 4 screws. 

8. The keyboard is firmly attached to the case by 10 screws (9 metal, 1 Nylon) 

9. The battery pack is firmly attached to the PCB. 

10. The two fuses FUS1 and FUS2 are not blown. 

11. The soldered points of the battery pack are without any damages and provide a 
safe connection. 

12. The 3 LEDs are firmly soldered to the keyboard PCB. 

13. The room temperature sensor is near the housing. 

14. The JP2 jumper (battery charging mode) is in the default position. 

15. The protection plastic angle for IC12 in the angle is present and not damaged. 

3.4. REPLACEMENT PROCEDURES 

3.4.1. General 

When handling sensitive static devices such as the mainboard of the spirolab the 
following precautions should be observed: 

Persons should be earthed by means of a wrist strap. 

Ground all electrical equipment, workbench etc. 

Caution: 

Removing SJ1 jumper will cause a loss of all data. After having removed one or more 
of the jumpers make sure to replace them properly when the repair is finished. 

Caution: 

Even if the unit is switched off the internal batteries are active and are supplying 
current so there are still voltages available, that can be easily short-circuited. To avoid 
unforeseen damaging through voltage, removal of all jumpers is recommended. 

Before removing the jumpers remember that removing the SJ1 jumper will cause a 
loss of all test data in memory. 

Also make sure that you do have left any tools inside and all screws including their 
washers are again fixed in the original place. 


Page 16 of 33 




MIR srl 


Spirolab / Spirolab II - Service Manual 


Note: Where not otherwise indicated, the reinstallation of a part is in reverse order of 
the removal procedure. 

After every repair a functional test must be performed. 

3.4.2. Cover 

Remove: Switch off the unit and close the display. Disconnect the battery charger 
cable from the device. Disconnect the turbine flowmeter. Turn the spirolab upside- 
down. Remove the screws with a 2.5mm Allen key. Turn the spirolab to its original 
position. 

Caution: 

Be careful not to damage the cables while opening the spirolab. 

Carefully lift the cover on the backside about 50 mm from the chassis. Then pull the 
cover about 50 mm towards you. In this way you create just enough space between 
the chassis and the cover to remove the cables. 

Disconnect the display backlight supply (CN 3) and the display cable (CN 4). 

Note: The keyboard cables (CN 8, CN 9) are not disconnectable. 

Turn the cover over the front edge, so that cover is lying next to the chassis. 

3.4.3. Printer 
Warning: 

The printer motor could be very hot! Do not touch! 

Removal: See para. 3.4.2 for removal of the cover. Remove the paper out of the 
printer. Remove the screws which fix the printer to the chassis. 

Disconnect the wiring which connects the printer to the mainboard from connectors 
CNB1, CNC and CNA1. 

Remove the printer. 

3.4.4. PCBs and components 

3.4.4.1. Keyboard 

Removal: See para. 3.4.2 for removal of the cover. Remove the screws which attach 
the keyboard to the cover. Disconnect the mainboard with a soldering iron and take 
off the keyboard. 

3.4.4.2. Complete PCB-unit (mainboard + keyboard) 

Removal: See para. 3.4.2 for removal of the cover. See para. 3.4.3 for removal of the 
printer. Remove the screws which attach the PCB to the chassis. Remove the screws 
which attach the keyboard to the cover. Lift it up on the battery side and after having 
reached an almost nearby vertical position, remove it. 


Page 17 of 33 



MIR srl 


Spirolab / Spirolab II - Service Manual 


3.4.4.3. Components 

Caution: 

If any solderings (especially on SMD-para.ts) are to be made, make sure, that your 
tools are suitable for such precision work. 

Removal: See para. 3.4.2 for removal of the cover. See para. 3.4.3 for removal of the 
printer. Remove the screws, which attach the PCB to the chassis. Remove the screws, 
which attach the keyboard to the cover. Lifting it on the battery’s side and after 
having reached a nearby vertical position, remove it. 

3.4.5. Membrane keyboard 

Removal: See para. 3.4.2 for removal of the cover. Remove the screws which attach 
the keyboard to the cover. Lift the keyboard and remove the membrane keyboard. 

3.4.6. Display 
Warning: 

In order to prevent unforeseen incidents, keep away from sensitive electronical 
devices. Feeding the display high voltages are present (max. 450 V), but at very 
low currents (max. 6 mA) so these represent no danger for healthy persons with 
experience and general knowledge of safety precautions. 

In exceptional cases these high voltage may interfere with sensitive electronical 
devices. 

Removal: See para. 3.4.2 for removal of the cover. Lift with moderate force the 
display first on the side where the supplementary cables leave the display’s housing, 
then at the other side. 

3.4.7. Battery pack 

The instrument is powered by rechargeable batteries (6 elements Ni-MH 1,2 V, 4000 
mAh). A 3,7 V lithium battery maintains the backup of the SRAM and RTC. 

Warning: 

The batteries may explode if defective or damaged or disposed of in fire. 

Do not short-circuit the batteries! 

Removal: See para. 3.4.2 for removal of the cover. See para. 3.4.4.2 for removal of the 
PCB. Disconnect the wiring which connects the battery to the mainboard from 
connectors CN6. Turn the mainboard over, detach the battery and remove it. 

Note: Always replace the battery with the same type of battery. 

The internal batteries must be disposed off according to the local regulations. 

3.4.8. Optoelectronic head 

The optoelectronic head is a complete assembly. Inside are no serviceable items. 
When defective, the complete head must be replaced. 


Page 18 of 33 



MIR srl 


Spirolab / Spirolab II - Service Manual 


3.4.9. Turbine 

Removal'. Pull the turbine/mouthpiece adaptor out of the turbine flowmeter. Pull the 
turbine out of the mouthpiece adaptor. 

3.5. ADJUSTMENT PROCEDURES 

3.5.1. Spirolab unit 

Note: The spirolab turbine measurement system is calibrated in the factory and does 
not require any adjustments or calibrations. 

3.5.2. Turbine flowmeter 

For proper working the flowmeter must be kept clean. For information about cleaning 
the flowmeter see the User’s Manual. 

Calibrating the flowmeter is only possible by the “turbine calibration” function of the 

spirolab. 

Please refer to the User’s Manual. 


Page 19 of 33 



MIR srl 


Spirolab / Spirolab II - Service Manual 


4. SPARE PARTS 

This chapter deals with placing orders for spare parts (see para. 6.1) as well as the 
listing of actual spare parts (appendix A, spare parts list). 

4.1. ORDERING 

Spare parts can only be ordered if the following data is supplied: 

- Part number 

- Description 

- Quantity 

- Shipping details (Incoterms) 

The required data is mentioned in appendix A. For the current parts programme 
please refer to the spare parts price list which is published every year. 

MIR only supplies parts mentioned in appendix A. Standard parts such as screws, 
nuts, cleaning agents and so on are not supplied by MIR, these parts must be 
purchased locally. 

Send your order to your contact person at the sales department of MIR. After 
receipt of an order you will receive a confirmation. Our minimum order value is 
250 Euro. Always check the ordering data. 

4.1.1. Ordering PCB's 

Printed circuit boards are only obtainable as spare part in their latest version. 
Consult the technical documentation for instructions of how to modify the latest 
version of the PCB for use in older versions of the equipment. 

4.1.2. Warranty claims 

Warranty claims must be provided with the MIR invoice number, the type of 
products and serial number of the equipment in question. The defective item must 
be returned back to MIR. The customer is responsible for the transportation and for 
all transport and customs charges for the delivery of the goods both to and from the 
service centre. 

4.2. RETURN SHIPMENTS 

- Any instrument or accessory returned must be accompanied by a clear and 
detailed explanation of the defect or problem found. 

- If units are to be returned to MIR then a written or verbal permission (RAN - 
Return Authorization Number) must be received before any instrument is returned 
to MIR. 

- The unit must be returned in its original packaging. 

- MIR reserves the right to modify the instrument if required, and a description of 
any modification made will be sent along with the returned goods. 


Page 20 of 33 




MIR srl 


Spirolab / Spirolab II - Service Manual 


Note: Any instalment of accessory returned must be accompanied by a clear and 
detailed explanation of the defect or problem found. 


Page 21 of 33 



MIR srl 


Spirolab / Spirolab II - Service Manual 


5. TROUBLESHOOTING 

Note: Where not indicated, voltages are measured versus the ground (i.e. the negative 
pole of the battery pack). 

Note: Where not indicated, voltages are measured while the unit is switched on. 

Note: Where not indicated, the values to be measured can vary by about 2%. 

Warning: 

Be careful not to short-circuit any contacts while measuring voltages. 

Batteries may explode if defective or damaged or disposed of in fire. 

Do not short-circuit the batteries! 

5.1. THE DEVICE DOES NOT SWITCH ON 

0 Check the voltage of the internal battery pack. 

- Open the spirolab (refer to para. 3.4.2) 

- Measure the voltage between the two poles of the battery pack, it must be 
between 6 V (low battery) and 8 V (full charged). 

0 Check the internal power supply. 

- The voltage on pin 3 of IC 29 must be 5 V. 

- The voltage on pin 3 of CN 9 for the ON LED must be approx. 1.8 V. 

0 If the battery charger is connected: 

Check the battery charger: 

- Make sure that the main power supply’s jack is correctly attached to the plug. 

- Check the cable for visible damage. 

- The voltage between inner and outer part of the jack must be 12 V. 

Check the internal power supply: 

- Open the spirolab (refer to para. 3.4.2). 

- Measure the voltage behind the fuse (FUS1) on the point indicated on the 
attached drawing. It must be approximately 12 V. 

0 Check the voltage on R1. It must be 5 V. 

0 Check the voltage on pins 2 and 3 of IC 24. They must be 0 V. 

0 Check the voltage on L6. It must be 5 V. 

5.2. THE LCD DOES NOT DISPLAY 
Warning: 

In order to prevent unforeseen incidents keep away from sensitive electronical 
devices. Feeding the display high voltage is present (max. 450 V) but at very low 
current (max. 6 mA), so that this represents no danger for healthy persons with 
experience and general knowledge of safety precautions. 

In exceptional cases these high voltage may interfere with sensitive electronical 
devices. 

0 Check the data connection 

- Open the spirolab (refer to para. 3.4.2). 

- Check the flat cable CN 4. 

- Check if the connector CN 4 is closed. 

- Check if the adhesive protection on CN 4 is attached. 

0 Check the mainboard’s function 

- Open the spirolab (refer to para. 3.4.2) 

- Switch on the device. 


Page 22 of 33 




MIR srl 


Spirolab / Spirolab II - Service Manual 


- Check the voltage between pin 1 and 4 of CN 3. This requires an oscilloscope. 
The amplitude peak to peak must be approximately 400 V at a frequency of 
approx. 60 kHz. 

0 Check the voltage on pin 7 of CN 4. This must be between 20 and 24 V, 
depending on the regulation of the display contrast. 

0 Check the voltage on pin 7 (RESET signal) of IC 34. It must be 5 V. 

0 Check the backup power supply (lithium battery): 

- Measure the voltage on pin 32 of IC 6. When the unit is switched off it must 
be 3.6 V equal to the voltage of the lithium battery BT1. When the unit is 
switched on it must be 5 V. 

- Check the voltage on pin 3 of IC 22. It must be 3.3 V. 

0 Check the voltage on pin 5 of CN 4. It must be 5 V. 

5.3. THE BATTERY CHARGING IS NOT WORKING CORRECTLY 

0 Check the battery charger 

- Make sure that the main power supply’s jack is correctly attached to the plug. 

- Check the cable for visible damage. 

- The voltage between the inner and outer parts of the jack must be 12 V. 

0 Check the internal power supply 

- Open the spirolab (refer to para. 3.4.2). 

- Measure the voltage behind the fuse (FUS1) on the point indicated on the 
attached drawing. It must be approximately 12 V. 

0 Check the thermistor 

- Open the spirolab (refer to para. 3.4.2) 

- Remove the thermistor cable from connector CN6. The resistance of the 
thermistor must be about 10 kO at room temperature (23°). 

- Replace the thermistor cable firmly to the connector CN 6. 

0 Check the red LED (CHARGE LED). 

- Open the spirolab (refer to para. 3.4.2) 

- Check the voltage on pin 4 of CN 9 (Flexstrip). It must be 1.8 V. 

5.4. THE LITHIUM BACKUP BATTERY INDICATES DISCHARGED 

0 Check the voltage of the lithium battery 

- Open the spirolab (refer to para. 3.4.2) 

- Check the voltage of the battery BT1 at SJ1 jumper, it must be 3.6 V. 

Warning: 

Batteries may explode if defective or damaged or disposed of in fire. 

Do not short-circuit the batteries! 

Caution: 

Removing the jumper at SJ 1 or the battery will cause a loss of ALL data in memory! 
Note: Batteries must be disposed off properly, according to the local regulations. 

5.5. THE PRINTER DOES NOT PRINT 

0 Make sure that suitable convenient thermal paper is inserted (112 mm width). 

0 Check the printer connections 

- Open the spirolab (refer to para. 3.4.2) 

- Remove the 4 screws that attach the printer to the housing. 

- All 3 flat cables must be firmly attached to CNA, CNB, CNC. 


Page 23 of 33 



MIR srl 


Spirolab / Spirolab II - Service Manual 


0 Check the voltage on R 77. 

- Open the spirolab (refer to para. 3.4.2) 

- Remove the printer (refer to para. 3.4.3) 

- Check the voltage on R 77 during printing. It must be 5 V. 

0 Check the voltage on the paper drive motor. 

- Open the spirolab (refer to para. 3.4.2) 

- Remove the printer (refer to para. 3.4.3) 

- Check the voltages on pins 2 and 9 of IC 9 and IC 10 while the printer is printing. 
They must be approx. 7 V depending on the battery level. 

5.6. THE PRINTER MAKES NOISES 

0 The printer is not firmly screwed to the body. 

- Open the spirolab (refer to para. 3.4.2) 

- Tighten the 4 screws using moderate force. 

5.7. THE DEVICE DOES NOT MEASURE AT ALL 

0 Check if any obstacle is blocking the free rotation of the turbine 
0 Check the cable for any visible damage. 

0 For information about maintenance cleaning and sterilization please refer to the 
relevant paragraph in the User’s Manual. 

5.8. THE DEVICE DOES NOT MEASURE CORRECTLY 


Inspiration 


0 Check if any obstacle is disturbing the free rotation of the turbine. 

0 For information about maintenance cleaning and sterilization please refer to the 

relevant paragraph in the User’s Manual. 

0 Check the turbine’s power supply 

- Open the spirolab (refer to para. 3.4.2) 

- Remove the printer (refer to para. 

3.4.3) 

- Press FVC to activate the turbine’s 
power supply 

- Check the voltage on pin 7 of CN2 
connector. It must be 5 V. This signal is 
0 V when FVC measurement is 
terminated. 

0 Check the turbine’s signal. This requires 

an oscilloscope. 

- Open the spirolab (refer to para. 3.4.2) 

- Picking up the signal before LI while 
blowing into the turbine, you see the 
original pulse signal at approx. 4V (see 
the picture on the right). 

- Behind LI, the signal must be more clearly at approx. 4V. On pin 12 of IC 14, 
you must see this signal inverted and squared at 5 V. 

- The same procedure is to be made at L2 and pin 10 of IC 14. 



m 





1 

; f 



m 


ifk 


ft 




« 

fr 

I: 



fr 

*> 










il 


J 



1? 

•krfrjJ 

rri 

1 












sWSffl 





a 


Exspiration 


5 ms 
2.00 





sgjs# j 








: !'<i 


Page 24 of 33 














































































MIR srl 


Spirolab / Spirolab II - Service Manual 


0 Check the conversion factors for BTPS: 


Temp. 

BTPS 

9°C 

1.164 

10°C 

1.160 

11°C 

1.155 

12°C 

1.151 

13°C 

1.147 

14°C 

1.142 

15°C 

1.138 

16°C 

1.133 

17°C 

1.129 

18°C 

1.124 

19°C 

1.120 


Temp. 

BTPS 

20°C 

1.115 

21°C 

1.111 

22°C 

1.106 

23°C 

1.101 

24°C 

1.097 

25°C 

1.092 

26°C 

1.087 

27°C 

1.082 

28°C 

1.078 

29°C 

1.073 

30°C 

1.068 


Temp. 

BTPS 

31°C 

1.063 

32°C 

1.058 

33°C 

1.053 

34°C 

1.048 

35°C 

1.043 

36°C 

1.038 

37°C 

1.033 

38°C 

1.028 

39°C 

1.023 

40°C 

1.018 

41°C 

1.013 


Note: The indicated values are calculated for a humidity of 50%. The more the 

humidity differs, the grater is the deviation of the calculated values. 

0 Check the accuracy with the calibration syringe. 

- Make several tests FYC at different in- and expiration speeds. 

- Check the measured volume. It may differ ± 3% or 50 ml, depending on which 
value is bigger (i.e. using a 3000-ml-syringe, the measured valued must be 
between 2910 and 3090 ml, using a 1000-ml-syringe, between 950 and 1050 
ml) 

0 Check the voltage on the ambient (room) temperature sensor IC 12 

- Open the spirolab (refer to par. 3.4.2) 

- The voltage on pin 4 of C 13 must be 100 mv/°C (i.e. 2,5 V for an ambient 
temperature of 25 °C). 

5.9. THE DATA COMMUNICATION VIA RS232 DOES NOT FUNCTION 


0 

0 


0 



Check the cable, if there are any visible damages. 

Check the RTS function for “line power supply” 

- Open the spirolab (refer to par. 3.4.2) 

- During the RS232 transmission measure the voltage between pin 7 and pin 5 
(ground of the connected device) of CN1. It must be “high level” (between +3 
and +12 V). 

Check data transfer while transmitting data from spirolab on the RS232 port (this 

requires an oscilloscope with memory) 

Note: Make sure, that you have connected the oscilloscope’s ground of the 

connected device (PC, printer, modem etc.) at pin 5 of CN1. 

- Press the 1 key (PC connection function) to start a process that sends the 
whole SRAM data via the RS232 port. This process endures about 10 minutes, 
but can be interrupted at any time by pressing [ ]. 

- Pick up the data stream at pin 2 of CN 1. You must see pulses at 9600 baud 
(9600 bit/sec) at a voltage in the range of RS232 standard (refer to 2.1.1.4). 


Page 25 of 33 










































MIR srl 


Spirolab / Spirolab II - Service Manual 


0 Start the application to update the internal software. The process must begin with 
the signal flow shown below. The voltage of the signals must be between High 
and Low level following the RS232 standard (refer to 2.1.1.4). 



Rx (PC) 


A: PC drives TX and DTR signals to a default state 

A-B: PC transmit 500 bytes continously and spirolab switches automatically on. 

C: PC drives the DTR signal Low 

D: spirolab responds transmitting a byte 

E: PC drives the DTR signal High 

F: PC drives the DTR signal Low 

G: spirolab responds transmitting a second byte 

H: PC transmits a command (byte) to start the flash erasing process. 

The same signals can be picked up behind the optocoupler: 

• TX of the PC on pin 4 of IC 14 

• DTR of the PC on pin 2 of IC 14 

• TX of the spirolab (RX of the PC) on Pin 6 of IC 14 

Note: For this measurement you must connect the oscilloscope’s ground to the 
spirolab’s ground (i.e. the negative pole of the battery pack) 

In this case, the signals are at logic level 0 / 5 V. 


Page 26 of 33 

































MIR srl 


Spirolab / Spirolab II - Service Manual 


5.10. INDEX OF COMPONENTS 

To simplify component identification they are marked in the attached drawing of the 
components layout of the Printed Circuit Board (PCB) MIR 009, rev. 3. 

The revision number of the PCB is indicated on the component side on the lower 
right, next to the battery pack). 

Comp.-ID Description.No. in drawing 


BT1 Lithium battery for data retention.8 

CN1 RS232 serial communication port.23 

CN2 Connector for the optoelectronic turbine head.1 

CN3 Connector for backlight supply of the LCD.18 

CN4 Connector for the display data cable.17 

CN6 Connector for the thermistor of the battery pack.13 

CN8 Flatcable for the keyboard.35 

CN9 Flatcable for the keyboard.34 

CNA Connector for supply, data transmission and printer 

head temperature.5 

CNB Connector for power supply of the printer motor.3 

CNC Connector for sensor that checks the paper.7 

FUS1 2 A battery charging controller fuse (on the mainboard).15 

FUS2 4 A mainboard fuse.12 

IC1 Microcontroller.29 

IC6 SRAM memory 4 Mbit.28 

IC9 Driver for the paper’s motor.2 

IC10 Driver for the paper’ s motor.4 

IC11 Optocoupler for printer head and motor power supply enable.6 

IC12 Ambient (room) temperature sensor.26 

IC13 Operational amplifier for ambient temperature sensor (IC12).25 

IC 14 Hex inverter with Schmitt trigger.21 

IC 22 Voltage regulator 3.3V output for display controller (IC21).31 

IC23 I/O decoder for display controller (IC21).33 

IC24 RTC.19 

IC29 Voltage regulator 5V output for device switch on.27 

IC34 RESET and data retention controller.9 

JP1 Jumper selector to program the EEPROM.20 

JP2 Jumper for adjusting the battery charging termination mode.14 

LI Inductor for turbine signal filter.24 

L2 Inductor for turbine signal filter.22 

L6 Inductor for the DC/AC converter of LCD backlight.16 

NEG. pole Negative pole of the NiMH battery pack.30 

POS. pole Positive pole of the NiMH battery pack.36 

R1 Resistor.32 

SJ1 Jumper for connecting the battery pack with the unit.10 

S J2 Jumper for connecting the Lithium battery with the unit.11 


Page 27 of 33 









































MIR srl 


Spirolab / Spirolab II - Service Manual 


APPENDIX A: SPARE PARTS LIST 


Product description 

Product 

code 

Model 


Axis for paper roll 

340832 

SL 1 /SL II 

1 

Battery pack 6 x 1.2 V, 4 Ah 
(rechargeable, internal) 

972301 

SL 1 /SL II 

1 

Lithium battery 3,6 V 

970232 

SL 1 / SL II 

1 

Battery charger 220V 

920665 

SL 1 / SL II 

1 

Battery charger 110V 

920675 

SL 1 / SL II 

1 

Buzzer BZ 1 

560020 

SL 1 / SL II 

1 

Carrying case (soft) 

672685 

SL 1 / SL II 

1 

Display (black/white) complete 
with flat connecting cable 

660591 

SL 1 

1 

Display (colour FSTN) complete 
with flat connecting cable 

660056 

SL II 

1 

Flat cable to connect display to 

PCB 

450100 

SL 1 / SL II 

1 

Fuse FI, 2 A, SMD 0805 

270464 

SL 1 / SL II 

1 

Fuse F2, 4 A, SMD 0805 

270468 

SL 1 / SL II 

1 

LED orange for PCB keyboard 

251900 

SL 1 / SL II 

1 

LED green for PCB keyboard 

251906 

SL 1 / SL II 

2 

Membrane keyboard 

221540 

SL II 

1 

Optoelectronic turbine holder with 
cable (without turbine flow meter) 

910012 

SL II 

1 

Paper roll for printer 

910350 

SL II 

20 

PCB keyboard without mainboard 

450055 

SL 1 / SL II 

1 

PCB mainboard without LCD 
display and without printer 

001040 

SL 1 / SL II 

1 

Plastic case (ABS) blue, complete 
including screws 

301040 

SL 1 / SL II 

1 

RS232 cable (spirolab to PC) 

671492 

SL 1 / SL II 

1 

RS232 port CN 1 

124556 

SL 1 / SL II 

1 

Rubber feet, set of 4 

301301 

SL II 

1 set 

Serial-Parallel printer converter 

910110 

SL 1 / SL II 

1 

Thermal printer 

660075 

SL 1 / SL II 

1 

T ransformer T1 

350001 

SL 1 / SL II 

1 

Turbine flow meter 

910000 

SL 1 / SL II 

1 


Page 28 of 33 





















































































































MIR srl 


Spirolab / Spirolab II - Service Manual 


APPENDIX B: SERVICE INFO'S (Product Change Notes) 

General 

The technical information in the service manuals of MIR srl is up to date at the date of 
issue. 

Necessary additional information (of any kind) will be provided in the following way: 

as supplements to the manual, or 

as revised editions (chapters) of the manual, or 

as Service Info (product change notes). 

When a supplement or a revision is made to the manual, a "document history file" is 
added to this chapter. This file will give information about the date of issue of 
supplements or revisions. 


Page 29 of 33 




MIR srl 


Spirolab / Spirolab II - Service Manual 


REVISION HISTORY 

First edition: Rev. 1 

Revisions: 

Chapter: Revision: 


June 2001 


Date of issue: 


page 


Page 30 of 33 




MIR srl 


Spirolab / Spirolab II - Service Manual 


Index 

1. INTRODUCTION 2 

1.1. INTRODUCTORY REMARKS 2 

1.1.1 .General Aspects.2 

1.1.2.Safety Aspects.2 

1.1.3. Legal Aspects.3 

1.1.4. Environmental Aspects.3 

1.2. DATA REGISTRATION 3 

1.2.1. Equipment and customer registration / Service file 3 

1.2.2. Configuration register 3 

1.2.3. Product documentation 4 

1.2.4. Installation 4 

1.2.5. Technical modifications 4 

1.2.6. Product evaluation 4 

1.2.7. Additional Information 5 

1.3. GENERAL 6 

1.3.1. Technical Data 6 

1.3.1.1. Battery charger 6 

1.3.1.2. Spirolab unit 6 

1.4. Standards Applied 7 

2. HARDWARE DESCRIPTION 8 

BLOCK DIAGRAM 8 

2.2. MAINBOARD MODULE 9 

2.2.1. Jumpers 9 

2.2.2. Charging controller for battery pack 9 

2.2.3. Room temperature sensor 11 

2.2.4. RS 232 communication port 11 

2.3. Display module 11 

2.4. Printer module 12 

2.5. Keyboard module 12 

2.6. Turbine flowmeter 12 

3. MAINTENANCE 15 

3.1. GENERAL 15 

3.2. TEST EQUIPMENT 15 

3.3. CHECKLIST 15 

3.3.1. Functional test 15 

3.3.1.1. Self Test 15 

3.3.1.2. Software version 16 

3.3.1.3. Hardware control 16 


Page 31 of 33 








MIR srl 


Spirolab / Spirolab II - Service Manual 


3.4. REPLACEMENT PROCEDURES 16 

3.4.1. General 16 

3.4.2. Cover 17 

3.4.3. Printer 17 

3.4.4. PCBs and components 17 

3.4.4.1. Keyboard 17 

3.4.4.2. Complete PCB-unit (mainboard + keyboard) 17 

3.4.4.3. Components 18 

3.4.5. Membrane keyboard 18 

3.4.6. Display 18 

3.4.7. Battery pack 18 

3.4.8. Optoelectronic head 18 

3.4.9. Turbine 19 

3.5. ADJUSTMENT PROCEDURES 19 

3.5.1. Spirolab unit 19 

3.5.2. Turbine flowmeter 19 

4. SPARE PARTS 20 

4.1. ORDERING 20 

4.1.1. Ordering PCB's 20 

4.1.2. Warranty claims 20 

4.2. Return shipments 20 

5. TROUBLESHOOTING 22 

5.1. The device does not switch on 22 

5.2. The LCD does not display 22 

5.3. The battery charging is not working correctly 23 

5.4. The lithium backup battery indicates discharged 23 

5.5. The printer does not print 23 

5.6. The printer makes noises 24 

5.7. The device does not measure at all 24 

5.8. The device does not measure correctly 24 

5.9. The data communication via RS232 does not function 25 

5.10. Index of components 27 

APPENDIX A: SPARE PARTS LIST 28 

APPENDIX B: SERVICE INFO'S (Product Change Notes) 29 

REVISION HISTORY 30 

COMPONENT LAYOUT 33 


Page 32 of 33