Skip to main content

Full text of "Exergen TemporalScanner TAT 5000 Operator’s Manual"

See other formats


English 


EXEREEN 

Temporal Scanner™ 


. 

5sJI^ 


-I- 

DH 

i“a 

_1 


O 




Accurate Temperature with a 
Gentle Forehead Scan 



x r/ 





Four Reasons to Standardize 
with Exergen Thermometers 

1. More than 70 published studies supporting accuracy 
from preemies to geriatrics, in all areas of care. 

2. Independent studies show costs are reduced by 90% 
compared to other thermometers. 

3. Rugged, reliable construction, protected by Lifetime 
Warranty. 

4. TAT-5000S Connected Models available on leading 
Vital Signs Monitors for EHR data integration. 















Important Safety Instructions 


READ ALL INSTRUCTIONS BEFORE USING 

Intended Use: The Exergen TemporalScanner is a handheld infrared thermometer used by medical 
professionals for the intermittent measurement of human body temperature of people of all ages, by 
scanning the forehead skin over the temporal artery. Intended users are physicians, nurses, and nursing 
assistants at all levels who normally provide patient care. The thermometer provides a peak temperature 
reading from plural readings during the step of scanning. Electronic circuitry processes the measured peak 
temperature to provide a temperature display based on a model of heat balance relative to a detected arterial 
temperature, the electronic circuitry computing an internal temperature of the body as a function of ambient 
temperature (Ta) and sensed surface temperature. Training materials that are supplementary to this 
instruction manual are available at www.exergen.com/s, and recommended for first time users. 

TAT-5000 Series thermometers are used by medical professionals in clinical environments. Such medical 
professionals include physicians, nurses, nurses’ aides, patient care technicians, and others who are trained 
to take the temperature of patients. Clinical environments include areas where medical professionals are 
providing medical services for patients, including hospitals, outpatient clinics, primary care offices, and other 
settings where temperature is taken as part of patient care. Clinical environments include Emergency 
Medical Services environments. 

Additionally, the TAT-5000 series thermometers are not for use aboard aircraft or near High Frequency 
Surgical Equipment or Radio Frequency shielded rooms, such as MRI (Magnetic Resonance Imaging) areas. 

When using the product, basic safety precautions should always be followed, including the following: 

• Use this product only for its intended use as described in this manual. 

• Do not take temperature over scar tissue, open sores, or abrasions. 

• The operating environmental temperature range for this product is 60°to 104°F (15.5°to 40°C). 

• Always store and transport this thermometer in a clean, dry place where it will not become 
excessively cold (-4°F/-20°C), or hot (122°F/50 q C). Relative humidity 93% maximum non-condensing. 
Atmospheric pressure 50 kPa to 106 kPa. 

• The thermometer is not shockproof. Do not drop it or expose it to electrical shocks. 

• Do not autoclave. Please note cleaning procedures in this manual. 

• Do not use this thermometer if it is not working properly, if it has been exposed to 
temperature extremes, damaged, been subject to electrical shocks or immersed in water. 

• There are no parts that you can service yourself except for the battery, which you should replace 
when low by following the instructions in this manual. For service, repair, or adjustments, return your 
thermometer to Exergen. Warning: No modification of this equipment is allowed. 

• Never drop or insert any object into any opening. 

• If your thermometer is not used regularly, remove the battery to prevent possible damage due 
to chemical leakage. 

• Follow the battery manufacturer’s recommendations or your hospital policy for the disposal of 
used batteries. 


©2018 Exergen Corporation 


page 2 




• Not suitable for use in the presence of flammable anaesthetic mixtures. 

• If the device fails to operate as described above, see the FAQ section of this manual. Additionally, 
ensure that you are not in the presence of electromagnetic disturbances. 

• If you have any additional questions regarding use or care of the thermometer, please 
seewww.exergen.com or call customer service at (617) 923-9900. 


/jk Type BF Applied Part indicates degree of patient protection against electrical shock. The product is 
#\ internally battery powered and electrically isolated from earth. 


WARNING: Use of this equipment adjacent to or stacked with other equipment (other than TAT-5000 

compatible patient monitors) should be avoided because it could result in improper operation. 

If such use is necessary, this equipment and the other equipment should be observed to verify 
that they are operating normally. 

WARNING: Use of accessories, transducers and cables other than those specified or provided by the 
manufacturer of this equipment could result in increased electromagnetic emissions or 
decreased electromagnetic immunity of this equipment and result in improper operation. 


WARNING: Portable RF communications equipment (including peripherals such as antenna cables and 
external antennas) should be used no closer than 30 cm (12 inches) to any part of the TAT- 
5000 thermometer, including cables specified by the manufacturer. Otherwise, degradation of 
the performance of this equipment could result. 


SAVE THESE INSTRUCTIONS. 


©2018 Exergen Corporation 


page 3 



Product Map of the Exergen TemporalSca/?/?erTAT-5000 



Probe Cone 


Probe Lens 



ON Button 

Automatic turn-off in 30 
seconds (when measuring in 
body temperature range, 
otherwise 5 seconds) 

LED Display Screen 


Battery Compartment Door-^ 
Compartment Door Screw 


I 

>s! 




Battery Compartment 
9-volt battery 

9-Volt Battery 


Introduction to Temporal Artery Thermometry 


Temporal artery thermometry (TAT) is a unique method of temperature 
assessment, using infrared technology to detect the heat naturally 
emitting from the skin surface. In addition, and of key importance, this 
method incorporates a patented arterial heat balance system to 
automatically account for the effects of ambient temperature on the skin. 

This method of temperature assessment has been shown to improve 
results and reduce costs by non-invasively measuring body temperature 
with a degree of clinical accuracy unachievable with any other thermometry 
method. The temporal scanner is a type BF device. 

Before Using, Familiarize Yourself with the Instrument 

• To Scan: Depress the red button. The instrument will continually scan for the highest temperature (peak) as 
long as the button is depressed. 

• Clicking: Each fast click indicates a rise to a higher temperature, similar to a radar detector. 

Slow clicking indicates that the instrument is still scanning, but not finding any higher temperature. 

• To Retain or Lock Reading: The reading will remain on the display for 30 seconds after button is released. 
If measuring room temperature, the temperature will remain on the display for only 5 seconds. 

• To Restart: Depress the button to restart. It is not necessary to wait until the display is clear, the 
thermometer will immediately begin a new scan each time the button is depressed. 

Alternate sites when temporal artery or behind ear are unavailable: 



• Femoral artery: slowly slide the probe across groin. 

• Lateral thoracic artery: slowly scan side-to-side in the area midway between the axilla and the nipple. 


©2018 Exergen Corporation 


page 4 





















2-Step Infant Temperature Measurement 



Step 1 

Place probe flush on center of 
forehead and depress button. Keeping 
button depressed, slowly slide probe 
mid-line across forehead to the hair 
line. 



Step 2 

Release button remove from head 
and read. 


How to improve the accuracy of your measurements on infants 



The preferred site is the 
temporal artery area. Unless 
visibly diaphoretic, one 
measurement here is typically 
all that is required. 


If the temporal artery is 
covered, then the area behind 
the ear, if exposed, can be an 
alternate site. 


Measure straight across the 
forehead and not down side of face. 
At mid-line, the temporal artery is 
about 2 mm below the surface, but 
can go deeply below the surface on 
the side of the face. 


Brush the hair aside if 
covering the area to be 
measured. 

Measurement site must 
be exposed. 


3-Step Adult Temperature Measurement 



Step 1 

Slide across forehead. 

Place probe flush on center of forehead 
and depress button. Keeping button 
depressed slowly slide probe mid-line 
across forehead to the hair line. 


Step 2 

Slide behind ear. 

Keeping button depressed, lift probe from 
forehead, touch behind ear halfway down 
the mastoid process and slide down to the 
soft depression behind the earlobe. 


Step 3 

Release button and read. 


How to improve the accuracy of your measurements on adults 




Measure only the up-side on a patient 
in a lateral position. 

The down-side will be insulated 
preventing the heat from dissipating, 
resulting in falsely high readings. 


Think of a sweatband. 

Measure straight across the forehead 
and not down the side of the face. 

At mid-line, the temporal artery is about 
2 mm below the surface, but can go 
deeply below the surface on the side of 
the face. 



Measure exposed skin. 


Brush the hair and bangs aside 
if covering the area to be 
measured. 


©2018 Exergen Corporation 


page 5 
















FAQs 

How does the temperature from a temporal scanner relate to core temperature? 

Temporal artery temperature is considered a core temperature because it has been demonstrated as accurate as the temperature measured by a 
pulmonary artery and esophageal catheter, and as accurate as a rectal temperature on a stable patient. Rule of thumb: Rectal temperature is 
about 1 °F (0.5°C) higher than an oral temperature and2°F (1 °C) higher than an axillary temperature. It will be easy to remember if you think of 
core temperature as a rectal temperature, and apply the same protocol you would use for a rectal temperature. 

If your thermometer is marked Arterial/Oral and has a serial number beginning with “O” (standard model start with “A”), it is programmed to 
compute the normal average cooling effect at the mouth, and automatically reduces the higher arterial temperature by that amount. This 
calibration allows the hospital to maintain existing protocols for fever workups based on oral temperature, and results in a reading consistent 
with the 98.6°F (37°C) mean normal oral temperature, in the range of 96.6 - 99.5°F (35.9 - 37.5°C). 

What should I do if I get an abnormally high or low reading, how do I confirm my reading? 

• Repeat the reading with the same Temporal Scanner; a correct reading will be reproducible. 

• Repeat the reading with another Temporal Scanner. Two Temporal Scanners with the same reading will confirm the reading. 

• Sequential readings on the same patient in rapid succession will cool the skin; it is best to wait about 30 seconds for the skin 
to recover from the cold probe. 


Possible causes of abnormal readings. 


Type of 

abnormal 

Temperature 

Possible cause 

Helpful hint 


Dirty Lens 

Clean lens of scanner every two weeks. 


Releasing the button before 
finished measuring 

Release the button after finished measuring. 


Measuring when an ice pack or wet 
compress is on the forehead 

Remove ice pack or wet compress, wait 2 minutes, and re¬ 
take temperature. 

Abnormally low 
Temperature 

Measuring a completely 
diaphoretic patient 

Complete diaphoresis includes diaphoresis of area behind the 
ear and suggests that the temperature is rapidly dropping. 

Use an alternative method of temperature measurement in 
these cases until the patient is dry and the temporal artery 
measurement can be repeated. 


Improperly scanning down the 
side of the face 

Scan straight across forehead. The temporal artery is closest 
to skin in that area. 

Abnormally high 
temperature 

Anything covering the area to be 
measured would insulate and prevent 
heat from dissipating, resulting in false 
high readings. 

Confirm measurement site has not recently been in contact 
with heat insulators such as hats, blankets, and hair. 

Scan the area not covered or wait about 30 seconds for the 
previously covered area to equilibrate to the environment. 


DISPLAY DIAGNOSTICS CHART 

The following chart summarizes the conditions that may occur while the 
TemporalScanneris in use, and the associated indications: 


Condition 

Display 

Range 

High Target 

HI 

>110 °F (43 °C) 

Low Target 

LO 

<61 °F (16 °C) 

High Ambient 

HI A 

>104 °F (40 °C) 

Low Ambient 

LOA 

<60 °F (16 °C) 

Low Battery 

bAtt 


Dead Battery 

blank display 


Processing Error 

Err 

Restart. Return to 
Exergen for repair if 
error message persists. 

Scanning 
(Normal Operation) 

SCAN 



©2018 Exergen Corporation 


page 6 























Care and Maintenance 


• Battery: A standard alkaline 9V battery provides approximately 15,000 readings.** To 
replace, loosen the single screw at the bottom of the instrument and remove the battery 
cover. Disconnect the old battery and replace with a new one in the same location. 

Replace the cover, and tighten the screw. Use only high quality alkaline batteries. 

• Handling: The Temporal Scanner is designed and built to industrial durability standards 
in order to provide long and trouble-free service. However, it is also a high precision 
optical instrument, and should be accorded the same degree of care in handling as you 
would provide other precision optical instruments, such as cameras or otoscopes. 

• Cleaning the case: The TemporalScannercase can be wiped down using a cloth 
dampened with 70% isopropyl alcohol. The industrial grade housing and design of the 
electronic components allow for completely safe cleaning with 70% isopropyl alcohol but 
should not be immersed in fluid or autoclaved. 

• Cleaning the sensor lens: With normal use, the only maintenance required is to keep the lens 
on the end of the probe clean. It is made of special mirror-like, silicon infrared-transmitting 
material. However, dirt, greasy films or moisture on the lens will interfere with the passage of 
infrared heat and affect the accuracy of the instrument. Regularly clean the lens with a cotton 
swab dampened with an alcohol wipe. Use only light force for cleaning, to avoid damaging the 
lens. Water can be used to remove any residual film left by the alcohol. Do not use bleach or 
other cleaning solutions on the sensor lens. Use 70% isopropyl alcohol. 




CLEANING 

INSTRUCTIONS 


• Calibration: Factory calibration data is installed via a computer which communicates with the TemporalScanner’s 
microprocessor. The instrument automatically self-calibrates each time it is turned on using this data, and will never 
require recalibration. If readings are not correct, the instrument should be returned for repair. See instructions of the 
return process. 


Instructions for Fahrenheit or Celsius Conversion 


The TemporalScannercan be used in either °F or °C. The TemporalScannerwill come preset based on your preference at the 
time of purchase. To convert from one scale to the other, the only tool necessary is a small screwdriver. 


For °F/°C Conversion: 

• Loosen single screw on bottom of case and remove battery cover. 

• Remove battery. 

• Locate the switch to the right of the battery (shown in the drawing), and 
with the tip of the screwdriver, slide up or down to the desired scale. 

• Remove the screwdriver. 

• Replace battery and cover. 

Repair 

If repair is required: 

• Contact Exergen at (617) 923-9900 or repairs@exergen.com for a Return Materials Authorization (RMA) Number. 

• Mark the RMA number on the outside of your package and packing slips. 

• Include a description of the fault if possible. 

• Send the instrument to: 

Exergen Corporation 
400 Pleasant Street 
Watertown, MA 02472 USA 

• Include the address the instrument should be returned to. 



©2018 Exergen Corporation 


page 7 





Guidance and Manufacturer’s Declaration - Electromagnetic Emissions 


The infrared forehead thermometer model TAT-5000 series is intended for use in the electromagnetic environment specified 
below.The user of the TAT-5000 series should assure that it is used in such an environment. 


Emissions test 

Compliance 

Electromagnetic environment-guidance 

RF emissions CISPR 11 

Group 1 

The TAT-5000 series thermometer uses no RF energy therefore any emissions 
are unlikely to cause any interference in nearby electronic equipment 

RF emissions CISPR 11 

Class B 

The TAT-5000 series thermometer is suitable for use by a healthcare 
professional in a typical health care environment. 

Harmonic emissions 

Not applicable 

Voltage fluctuations 

Not applicable 


Guidance and Manufacturer’s Declaration - Electromagnetic Immunity 

The TAT-5000 series thermometer is intended for use in the electromagnetic environment specified below. The user of the TAT- 5000 
series should assure that it is used in such an environment. 

Immunity test 

IEC 60601 test 
level 

Compliance 

level 

Electromagnetic environment-guidance 




Portable and mobile RF communications equipment should be used 
no closer to any part of the TAT-5000 series including cables if 
applicable, than the recommended separation distance calculated 
from the equation applicable to the frequency of the transmitter. 

Conducted RF IEC 
61000-4-3 

3Vrms 

150 kHz to 80 MHz 

Not applicable 

Recommended separation distance 

d=1,2*P 1/2 

d=1,2*P 1/2 80 MHz to 800MHz 
d=1,2*P 1/2 800MHz to 2,5 GHz 

Radiated RF IEC 
61000-4-3 

3V/m 

80 MHz to 2,5 GHz 

3V/m 

Where P is the maximum output power rating of the transmitter 
in watts (W) according to the transmitter manufacturer and d is 
the recommended separation distance in meters (m). 

Field strength from the fixer RF transmitters, as determined by an 
electromagnetic site survey, a. should be less than the compliance 
level in each frequency range and b. interference may occur in the 
vicinity of equipment with the following symbol: 




(«i») 

Note 1 At 80MHz and 800MHz, the higher range applies. 

Note 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from 
structures, objects, and people. 

a. Field strengths from fixed transmitter, such as base stations for radio (cellular/cordless) telephones and land mobile radios, 
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the 
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the 
measured field strengths in the location in which the TAT-5000 series thermometer is used exceeds the applicable RF 
compliance level above, the TAT-5000 series thermometer should be observed to verify normal operation. If abnormal 
performance is observed, additional measures may be necessary, such as reorienting or relocating the TAT-5000. 

b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m. 

c. Portable and mobile RF communications equipment can affect performance. 

d. ESD compliance has been verified by testing. Intertek Report Number R11898679. 


©2018 Exergen Corporation 


page 8 





















Guidance and Manufacturer’s Declaration - Electromagnetic Immunity (cont) 

The TAT-5000 series thermometer is intended for use in the electromagnetic environment specified below. The user of the TAT-5000 
series should assure that it is used in such an environment. 

Immunity test 

IEC 60601 test level 

Compliance level 

Electromagnetic environment-guidance 

Electrostatic 

discharge 

(ESD) IEC61000-4-2 

6kV contact 8kV air 

6kV contact 8kV air 

Floors should be wood, concrete or ceramic tile. If 
floors are covered with synthetic material, the 
relative humidity should be at least 30%. 

Electrical fast 
transient/ 
burst 

IEC 61000-4-4 

2kV for power supply lines 
IkV for input output lines 

Not applicable 

Mains power quality should be that of a typical 
health care environment. 

Surge 

IEC 61000-4-5 

IkV line(s) to line(s) 

2kVline(s) to earth 

Not applicable 

Mains power quality should be that of a typical 
health care environment. 

Interruptions and 
voltage variations on 
power supply Input 
lines 

IEC 61000-4-11 

<5% UT 

(>95% dip in UT) for 0,5 cycle 

40% UT 

(60% dip in UT) for 5 cycles 

70% UT 

(30% dip in UT) for 25 cycles 

< 5% UT 

(>95% dip in UT) for5 sec. 

Not applicable 

Mains power is not applicable. The TAT-5000 
series is powered by battery and battery only. 

Power frequency 
(50/60 Hz) magnetic 
field 

IEC 61000-4-8 

3A/m 

3A/m 

Power frequency magnetic fields should be at the 
level characteristic of a typical location in a typical 
health care environment. 

Note UT is the a.c. mains voltage prior to the application of the test level 


Recommended separation distances between portable and mobile RF communication equipment and the TAT-5000 Series 

The TAT-5000 series forehead thermometer is intended for use in an electromagnetic environment in which radiated RF disturbances 
are controlled or the user of the TAT-5000 series thermometer can help prevent electromagnetic interference by maintaining aminimum 
distance between portable and mobile RF communications equipment (transmitters) and the TAT-5000 series thermometer as 
recommended below, according to the maximum output power of the communications equipment. 

Rated maximum output power 
of transmitter 
(W) 

Separation distance according to frequency of transmitter m 

150 KHz to 80 MHz 
d=1,2 P 1/2 

80 MHz to 800 MHz 
d=1,2 P 1/2 

800 MHz to 2,5 GHz 
d=2,3 P 1/2 

0,01 

0,12 

0,12 

0,23 

0,1 

0,38 

0,38 

0,73 

1 

1,2 

1,2 

2,3 

10 

3,8 

3,8 

7,3 

100 

12 

12 

23 

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can 
be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of 
thetransmitter in watts (W) according to the transmitter manufacturer. 

Note 1 At 80 MHz and 800 MHz the separation distance for the higher frequency range applies. 

Note 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from 
structures, objects and people. 


©2018 Exergen Corporation 


page 9 



























Accessories 


Disposable Cover Options 

(Model Illustrated: TAT-5000) 



No Cover No Cover 

Terminal Alcohol Wipe 

Cleaning at Between Patients 
Patient 


% ^ 

$ v. ML 



Probe Cap Full Sheath 

Covers Entire Covers Entire 

Probe Instrument 


Model TAT-5000 Options 


Accessories 

1. Instrument Holder (TAT-5000 not 
included) 

PN 134201 

2. Disposable Caps 
PN 134203 

3. Security Cables 

8 ft coiled cable - Latex free - PN 
124311 

6 ft. vinyl-covered steel - PN 134302 
8 ft. vinyl-covered steel - PN 134030 

4. Keyless Self-Locking Wall Mount 
PN 134305 

5. Keyless Self-Locking Wall Mount 
(shown with resposable cap 
dispenser) 

PN 134306 



The TAT-5000 can be used with either disposable caps or full sheath. 

Can be used without disposables if terminally cleaned or probe head disinfectant wipe between patients. Disposables are single 
patient use and can be re-used on the same patient. 


©2018 Exergen Corporation 


page 10 



















Specification 

TAT-5000 

Clinical Accuracy 

± 0.2°F or 0.1 °C Per ASTM E1112 

Temperature Range 

61 to 110°F (16 to 43°C) 

Arterial Heat Balance Range for 
Body Temperature* 

94 to 110°F (34.5 to 43°C) 

Operating Environment 

60 to 104° F (16 to 40 C C) 

Resolution 

0.1 °F orC 

Response Time 

~ 0.04 seconds 

Battery Life 

15,000 readings** 

Time Displayed on Screen 

30 seconds 

Size 

2.0” x 8.0” x 1.25” 

(5 cm x 20 cm x 3 cm) 

Weight 

7.9 oz. (223 g) 

Display Type and Size 

Large bright LED’s 

Construction Method 

• Industrial duty impact resistant casing 

• Hermetically sealed sensing system 

• Stainless steel probe 


Complies with EN 60601-1 3rd edition 


*Automatically applied when temperature is within normal body temperature range, otherwise reads surface temperature. 

** Approximate number of readings when scanning for 5 seconds and reading the temperature display for 3 seconds before turning thermometer off. 




Symbol for Manufacturer 

X 

Do not throw this device away in the trash, 
contact Exergen Corp. for disposal and 
recycling instructions. 

A 

Caution, Consult Accompanying 

Documents 

IPX0 

Ordinary Equipment 


m 

i 

Consult Instructions for Use 

o 

"On" (only for part of Equipment) 


ft 


Degree of Protection Against Electrical 
Shock 

Type BF Applied Part, Battery Operated 

E466615 

MEDICAL ELECTRICAL EQUIPMENT 
ANSI/AAMI/ES60601 -1: 2005/(R)2012 

3rd Edition including Amendment 1; 
CAN/CSA-C22.2 No. 60601.1: 2014; 

IEC 60601-1-6; ISO 80601-2-56: 

Particular Requirements 

For Basic Safety and Essential 
Performance of Clinical Thermometers 

For Body Temperature Measurement 


Exergen TemporalScanner Temporal Artery Thermometer 

Changing the Way the World TaHes Temperature 



EXERGEN CORPORATION • 400 PLEASANT STREET • WATERTOWN, MA 02472 • PH (617) 923-9900 


www.exergen.com 


©2018 Exergen Corporation 


page 11 





























http://www.exergen.com/medical/product/accessories.html 


Changing the Way the World Tahes Temperature 


©2018 Exergen Corporation 


page 12 


Document PN 818528 Rev 10